Pentacam系统4种模式测量角膜曲率的一致性研究
陈思萍, 陈茹茹, 黄锦海, 陈岩, 邓军
温州医科大学附属眼视光医院杭州院区眼视光专科
摘要:
目的采用Pentacam系统4种模式测量近视成人角膜曲率,评估其一致性,为临床使用提供参考。方法选取近视志愿受试者64人(64眼),应用Pentacam系统4种模式(分别是3D25、3D50、CF50、CF100)测量角膜3mm直径范围水平中央曲率K1、角膜3mm直径范围垂直中央曲率K2。4种模式测量的K1、K2比较采用重复测量方差分析,采用Bonferroni法两两比较模式间的测量差异,测量结果的一致性研究采用Bland-Altman分析。结果Pentacam系统4种模式测量的受试者K1分别为(42.86±1.49)、(42.86±1.49)、(42.88±1.50)、(42.86±1.49)D,K2分别为(44.00±1.74)、(43.99±1.74)、(43.99±1.76)、(43.99±1.73)D,4种模式测量的K1、K2比较均无统计学差异(均P>0.05),两两比较亦均无统计学差异(均P>0.05)。4种模式测量的K1、K2具有很好的一致性。结论Pentacam系统4种模式测量近视成人角膜曲率具有良好的一致性,在测量角膜中央曲率时可以替换使用。
关键词:  Pentacam 角膜地形图 一致性 测量模式
DOI:
分类号:
基金项目:温州市科技计划项目
Consistency evaluation of 4 different scan modes by Pentacam system for corneal curvature measurement in myopic eyes China
CHEN Siping, CHEN Ruru, HUANG Jinhai, CHEN Yan, DENG Jun
the Eye Hospital of Wenzhou Medical University at Hangzhou
Abstract:
Objective To evaluate the consistency of 4 different scan modes by Pentacam system for corneal curvature measurement in myopic eyes. Methods Central corneal curvature [horizontal (K1) and vertical (K2)] of 64 eyes were measured by Pentacam in 4 different scan modes (3D25, 3D50, CF50 and CF100).The measurements of 4 modes were compared with repeated-measure analysis of variance and Bonferroni multiple comparison correction.Bland-Altman analysis was used to evaluate the consistency of 4 modes. Results The mean K1 obtained by 4 modes was (42.86±1.49), (42.86±1.49), (42.88± 1.50) and (42.86±1.49)D, respectively.And the mean K2 was (44.00±1.74), (43.99±1.74), (43.99±1.76) and (43.99±1.73)D, respectively.There were no significant differences among the results of these 4 modes in K1 and K2 (both P >0.05).The Bland-Altman plot analysis showed that the 4 modes had comparable results for K1 and K2.For the maximum absolute values of 95% limits of agreement (CI) were 0.37 and 0.41D, respectively. Conclusion The 4 scan modes of Pentacam system demon- strate high consistency for K1 and K2 measurement in myopic eyes, indicating that 4 scan modes can be used alternatively for the measurements of central corneal curvature.
Key words:  Pentacam Corneal curvature Agreement Scan modes